Application Range
- Adult/Pediatric/Neonatal/Medicine/Surgery/Operating Room/ICU/CCU
Display
- 12.1”real color TFT screen
Environment
- Operating environment:
Temperature::0~40℃
Humidity:15%~80%
Pressure:86.0kPa~106.0kPa
Transport and Storage environment:
Temperature:-20~50℃
Humidity:10%~90%
Pressure:86.0kPa~106.0kPa
Power Requirements
- AC:100~240V,50Hz/60Hz
- DC:Built-in rechargeable battery
- Battery:11.1V 240wh Li-ion battery.2h operating after full charge.5min operating after low battery alarm
Date Storage
- Trend diagram/table:720h
- NIBP review:10000 events
- Wave review:12h
- Alarm review:200 alarm events
- Support drug concentration titration analysis
ECG
- 5 lead:RA,LA,LL,RL,V
- Lead mode:Ⅰ,Ⅱ,Ⅲ,AVR,AVL,AVF,V
- Gain:*1.25mm/mV,2.5mm/mV,5.0mm/mV,10mm/mV,20mm/mV(error≤5%)
- Sweep speed:6.25mm/s,12.5mm/s,25mm/s,50mm/s(error≤±10%)
- Bandwidth:surgery1~25Hz
- Monitor:0.5~40Hz
- Diagnostic 0.05~130Hz
- Sensitivity:>200μVP-P
- CMRR:≥112dB
- Differential Input Impedace:>5M
- ohm Electrode Offset Potential:±300mV
- Polarization Voltage:≥5M
- Calibration Voltage:1mV±5%
ST Segment
- St Segment Range:-2.0mV~﹢2.0mV
- Accuracy:0.02mV
NIBP
- Method:Oscillometry
- Measure mode:Manual,Auto,STAT
- Measure interval in auto mode:1~5/10/15/30/60/90/120/180/240/480min
- Measure period in STAT mode:5min
- Pulse rate range:40~240bpm
- Measure&alarm range:Adult
- SYS 40~280mmHg
- DIA 10~220mmHg
- MEAN 20~240mmHg
Pediatric
- SYS 40~220mmHg
- DIA 10~160mmHg
- MEAN 20~170mmHg
Neonatal
- SYS 40~135mmHg
- DIA 10~100mmHg
- MEAN 20~110mmHg
- Static pressure measurement
- Range:0~300mmHg
- Accuracy:±3mmHg
- Resolution:1mmHg accuracy:Max.Mean Error:±5mmHg,Max.Standard Deviation:8mmHg
- Overpressure protection:Adult 300mmHg,Pediatric 240mmHg,Neonatal 150mmHg
SpO2
- Measurement range:0~100%
- Resolution:1%
- Accruacy:70~100%±2%,40%~69%±3%
Pulse Rate
- Measure/alarm range:20~300bpm
- Resolution:1bpm
- Accuracy:±3bpm
RESP
- Method:impedance between RA-LL
- Resp impedance range:0.3~3Ω
- Base impedance range:200Ω~4000Ω
- Bandwidth:0.3~2.5Hz
- Gain:*0.25,*0.5,*1,*2,*4
- Resp rate:
- Measurement Range
- Adult 0~120BrPM
- Neonatal/Pediatric 0~150BrPM
- Resolution 1BrPM
- Accuracy:7~150BrPM(±2BrPM),0~6Brpm(unspecified)
- Apnea Alarm:10~40s
TEMP
- Channel 2
- Measure & alarm range:0~50℃
- Resolution:0.1℃
- Accuracy(no sensor):±0.1℃(0~50℃)
Standard accessories
- NIBP cuff & tube
- ECG cable & electrodes
- SpO2 sensor
- TEMP probe
- Lithium-ion battery
- Power cable
- Operator’s manual
Optional accessories
- IBP
- CO2
- Nellcor SpO2